Health Products

Negative Pressure Wound Therapy Sved Device

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Negative Pressure Wound Therapy Sved Device
Issue: Medical devices - Revised instructions for use
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
Negative Pressure Wound Therapy Sved DeviceAll lots.6701132BL, 6701132BL-DE, 6701132, 6701132BL-A

Issue

Cardinal Health is initiating an urgent medical device correction to update the clinician and patient user manuals for the Negative Pressure Wound Therapy Sved Device to add pertinent information related to periodic preventative maintenance for this device. The current three year service life stated in the clinician and patient user manuals has been augmented to include a requirement to service the units every 3 years or once the unit has reached 75 days of runtime, whichever comes first.

Recall start date: Apr 19, 2022

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime