· Published by Health Canada · Official notice
| Affected Products | Lot or Serial Number | Model or catalog number |
|---|---|---|
| Intelepacs | Version 4-10-1 and above | INTELEPACS |
There is a possibility that a patient's final report sent out of Intelepacs contains the preliminary text instead of the final report text. This issue can manifest itself only: - In specific pacs-driven workflow configurations (e.g. multi-method reporting with transcriptionist worklist) where the signing step is handled by Intelepacs, - Specifically when a network connectivity problem occurs between the server the user is connected to and the primary RIS gateway server (rig) of the system, and - When the rig servers are not fully synchronized.
Recall start date: Apr 25, 2022