Health Products

Ih-Qc 3

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Haematology
Product: Ih-Qc 3
Issue: Medical devices - Increased risk or new adverse events
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
Ih-Qc 308730 72 1 (SAP 709953 72 1)009323

Issue

Following customers' feedback, we have been able to confirm that the results obtained with this Ih-Qc 3 batch may not be as expected. Different issues were observed which may cause Qc failed results: a. Positive reactions may be observed in Direct Antiglobulin Tests (DAT) b. Weaker or in worse case false negative reactions may occur with Rh:4 cells in Indirect Antiglobulin Test (IAT) and in 2-stage papain technique c. When testing the impacted batch of Ih-Qc3 with Id-cards containing Anti-C, a double population with Anti-C (Rh:4) can be observed

Recall start date: Apr 25, 2022

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime