Health Products

Somatom Force

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Somatom Force
Issue: Medical devices - Software defect
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected ProductsLot or Serial NumberModel or catalog number
Somatom Force76330, 76458, 7644810742326

Issue

During the transmission of the patient table position information, sporadic interference signals may occur due to an error within the patient table position encoder. This problem can be detected by the system and can result in a scan abort. Sporadic aborts of perfusion scans related to this issue have been reported to us, which can occur during the patient table's back and forth movement when using the adaptive 4D spiral. This update instruction provides the revision 01 of the position sensor for the table top of affected PHS-5 and MPT-4 patient tables to prevent sporadic signal glitches at the zigzag movement of perfusion scans.

Recall start date: Apr 28, 2022

Get told about the next health product recall.