Health Products

ABL800 Flex System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: ABL800 Flex System
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturerABL800 Flex System393-801
393-800

Issue

Radiometer has become aware of a potential issue with abl800 basic and abl8xx flex analyzers with software versions below v6.19 mr2 and configured to report cna+. We have received sporadic incidents of positive and negative biases for cna+, that potentially could lead to serious health consequences. The issue may show when using specific micro measuring modes. The issue is caused by a combination of software and analyzer specific hardware related to sample transport.

Recall start date: May 10, 2022