Health Products

Bond Wash Solution

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Pathology
Product: Bond Wash Solution
Issue: Medical devices - Device compatibility
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Model or Catalogue numberLot or Serial numberAffected products
AR9590BATCH: BW102621Bond Wash Solution

Issue

Leica Biosystems recently received a complaint for cloudy bond wash, batch bw102621. The IFU states that turbidity of the solution indicates contamination and/or instability of the diluted wash solution. The functional testing was rerun and passed, however, the retain unit at working strength also exhibited cloudiness. This recall is to inform customers of the potential for this issue and to recommend that if the product has been used, the facility should determine whether a review of past results using the affected lot is clinically warranted and discard remaining unused units of this lot.

Recall start date: May 16, 2022

Get told about the next health product recall.