Health Products

Ortho Connect

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Ortho Connect
Issue: Medical devices - Software defect
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Health professionals, Hospitals
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Not applicableOrtho Connect6904410
6904411
6904412
6913355

Issue

An end user site attempted to perform QC on a single test using multiple QC kits. The QC module was not designed to support testing of multiple QC kits for a single test, and it is required that all QC samples used to validate a test are contained in a single QC kit in Ortho Connect™. If a test is processed across multiple QC kits, the QC status will be based on the results of only a single QC kit - if a single QC kit has a passing QC status, the status will be considered as passing for the test. This is the same failure mode that was previously communicated for Optix Reader regarding the quality control (QC) status displayed when using multiple QC kits.

Recall start date: May 4, 2022

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime