· Published by Health Canada · Official notice
| Model or Catalogue number | Lot or Serial number | Affected products |
|---|---|---|
| 1012083-120 1012084-120 1012087-120 1012087-150 1012082-120 1012082-150 1012085-120 1012086-120 1012085-150 1012086-150 1012088-120 1012089-120 1012088-150 1012083-150 1012084-150 1012089-150 | More than 10 numbers, contact manufacturer. | Absolute Pro Ll .035 Peripheral Self-Expanding Stent System |
Abbott has confirmed reports of mechanical locking, stent deployment, and partial stent deployment failures, resulting from unintended excessive force used to deploy the stent. To reduce occurrences of deployment failures and associated outcomes, Abbott is notifying all users of the potential causes and risks. This action does not affect patients having successfully undergone procedures using these devices.
Recall start date: May 24, 2022