Health Products

Eea Autosuture Circular Stapler With Dst Series

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General and plastic surgery
Product: Eea Autosuture Circular Stapler With Dst Series
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.Eea Autosuture Circular Stapler With Dst SeriesEEA25
EEAXL25
EEA2535
EEAXL2535

Issue

Notification to inform customers regarding potential for the staple guide to not be securely attached to the instrument. A staple guide not attached to the instrument could cause the component to disengage and if disengaged, could allow the device to transect tissue without forming staples. This could result in delay of treatment, extended hospital stay, unspecified tissue injury, unintended radiation exposure, unexpected medical intervention, foreign body in patient, failure to anastomose, and hemorrhage.

Recall start date: May 16, 2022

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime