· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| 64954T 64952T | Inteliq Specialty Immunoassay Control | 12008278 12008276 |
Bio-Rad laboratories would like to inform you that our Inteliq specialty immunoassay control (catalogue number 12008276, lot 64954t) and (catalogue number 12008278, lot 64954t) shows a stability issue. We have observed that PTH (intact) may not meet the thawed unopened, refrigerator off-board and instrument on-board storage stability claims as printed in the package insert.
Recall start date: May 24, 2022