Health Products

Inteliq Specialty Immunoassay Control

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Inteliq Specialty Immunoassay Control
Issue: Medical devices
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
64954T
64952T
Inteliq Specialty Immunoassay Control12008278
12008276

Issue

Bio-Rad laboratories would like to inform you that our Inteliq specialty immunoassay control (catalogue number 12008276, lot 64954t) and (catalogue number 12008278, lot 64954t) shows a stability issue. We have observed that PTH (intact) may not meet the thawed unopened, refrigerator off-board and instrument on-board storage stability claims as printed in the package insert.

Recall start date: May 24, 2022