· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| OS22871PLNBCX0004 OS22612PLNBCX0004 OS22873PLNBCX0004 OS22201PLNBCX0004 OS22334PLNBCX0004 OS23086PLNBCX0004 OS23189PLNBCX0004 OS24421PLNBCX0004 OS22303PLNBCX0004 | 1 Series Phototherapy Device | 807OS0004CX6 |
Software update required. a change was made to v3.00 (released june 30, 2021) to update bit11 in HST file flag. This was done for compatibility with clearlink website- discovered 0inch/9inch inconsistency when testing for Arcutis clinical trial data migration. Release versions prior to v3.00 did not have this functional issue.
Recall start date: June 6, 2022