Health Products

1 Series Phototherapy Device

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General and plastic surgery
Product: 1 Series Phototherapy Device
Issue: Medical devices - Performance
What to do: MANUFACTURER HAS AUTHORIZED USERS TO RETURN THE AFFECTED DEVICES FOR A SOFTWARE UPDATE.
Audience: General public
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
OS22871PLNBCX0004
OS22612PLNBCX0004
OS22873PLNBCX0004
OS22201PLNBCX0004
OS22334PLNBCX0004
OS23086PLNBCX0004
OS23189PLNBCX0004
OS24421PLNBCX0004
OS22303PLNBCX0004
1 Series Phototherapy Device807OS0004CX6

Issue

Software update required. a change was made to v3.00 (released june 30, 2021) to update bit11 in HST file flag. This was done for compatibility with clearlink website- discovered 0inch/9inch inconsistency when testing for Arcutis clinical trial data migration. Release versions prior to v3.00 did not have this functional issue.

Recall start date: June 6, 2022

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime