Health Products

Soltive Superpulsed Laser Ball Tip Fiber Single Use

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Soltive Superpulsed Laser Ball Tip Fiber Single Use
Issue: Medical devices - Labelling and packaging
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
KR149856TFL-FBX150BSSoltive Superpulsed Laser Ball Tip Fiber Single Use

Issue

Olympus has received a limited number of complaints that the laser fiber size on the label did not match the product's size. These complaints were identified by users during preparation for use. Olympus investigations confirmed that the model and size of laser fibers may be mismatched between the product's carton (box of 5), immediate pouch packaging and/or the device. One lot number was sold in Canada.

Recall start date: May 26, 2022

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime