Health Products

Dimension Exl System - Cardiac Troponin-I (CTnI) Assay

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Dimension Exl System - Cardiac Troponin-I (CTnI) Assay
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
GA2323Dimension Exl System - Cardiac Troponin-I (CTnI) AssayRF421C

Issue

Siemens Healthcare Diagnostics Inc. has confirmed that Dimension Cardiac Troponin I (lTni and CTnI) lots (ea2314 and ga2323) shows a positive bias with lithium heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/ml. Troponin values as high as 0.20 ng/ml have been observed with lithium heparin plasma that are expected to be within the 99th percentile. Serum samples collected from normal, healthy individuals whose results fall within the reference interval or below the 99th percentile of 0.07 ng/ml were not impacted by this issue.

Recall start date: May 24, 2022

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime