· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| 61658UQ01 59779UQ05 60229UQ07 61011UQ10 61349UQ01 60540UQ09 59561UQ04 | Alinity C Hemoglobin A1C Reagent Kit | 08P4320 |
| 60535UQ08 59588UQ04 61066UQ11 59868UQ05 60237UQ08 60698UQ09 60880UQ10 61246UQ01 60138UQ07 61250UQ01 59900UQ06 | Architect Hemoglobin A1C Reagent | 4P52-20 |
This recall is to inform of a potential for falsely elevated hemoglobin %A1C and hemoglobin A1C patient results to be generated using the architect hemoglobin A1C assay when analyzing whole blood or hemolysate samples with poor sample integrity, due to micro-clots and particulate matter, associated with low total hemoglobin values. to address this issue, a revision to the total hemoglobin low linearity value for the architect hemoglobin a1c whole blood and hemolysate applications has been implemented.
Recall start date: June 15, 2022