· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| All lots. | Cholestech Ldx Triglycerides Test | 10-991 10-989 |
| All lots. | Cholestech Ldx Glucose Test | 10-991 10-990 |
| All lots. | Cholestech Ldx Total Cholesterol Test | 10-987 10-991 10-990 10-989 |
| All lots. | Cholestech Ldx Hdl Cholesterol Test | 10-987 10-991 10-990 10-989 |
Alere San Diego has identified that bilirubin interference level listed in the package insert is incorrect. Concentrations less than 5.0 mg/dl can impact the accuracy of triglyceride (TRG) and High-density lipoproteins (HDL). Bilirubin levels of 2.0 mg/dl (conjugated) or 1.6 mg/dl(unconjugated) were found to be the correct concentrations. this change impacts all lots produced since product launch.
Recall start date: June 13, 2022