Merlin Patient Care System - Programmer Software

Issue

Abbott is notifying customers of the potential for programmer software (Merlin pcs) and remote monitoring software applications (Merlin.net) to display overestimated predicted battery longevity for certain pacemakers. Pacemaker/battery functionality, therapy delivery, and longevity remain normal and within specifications. Voltage measurements and ERI (elective replacement indicator), which is based on direct voltage measurement, remain accurate.

Recall start date: June 16, 2022