| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| More than 10 numbers, contact manufacturer. | D015451NK D015251NK D015551NK D015351NK D015151NK | Needle Kit For Powered Driver |
| More than 10 numbers, contact manufacturer. | D001001 | Intraosseous Powered Driver |
| More than 10 numbers, contact manufacturer. | D015351MK D015451MK D015151MK D015551MK D015251MK | Manual Driver Needle Kit |
1. Difficulty separating stylet from needle, which could result in inadvertent removal of the entire needle assembly or inability to remove the stylet from an indwelling intraosseous needle, both resulting in functional loss of intraosseous access. 2. Needle safety mechanism may not deploy as the stylet is removed from the intraosseous needle post placement. 3. Metal discs sticking in power driver, rendering the driver unable to be used
Recall start date: June 20, 2022