· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| software v1.7.0 or higher | Teg 6s | 07-690 |
Haemonetics corporation internal software test team identified test results printed using a thermal printer attached to the Teg 6s instrument, in non-english languages, exhibited text errors. The error is only applicable when using software version 1.7.0 or higher when testing platelet mapping(plm) adp&aa cartridges and printing in languages other than English.
Recall start date: June 23, 2022