Health Products

Pro+ Mattress

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Pro+ Mattress
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Not applicable.Pro+ MattressP7923A01
Not applicable.Pro+ MattressP3255A02
Not applicable.Pro+ MattressP7924A01
Not applicable.Pro+ MattressP3255A01
Not applicable.Pro+ MattressP7924A03
Not applicable.Pro+ MattressP7923A04
P7923A03

Issue

The Pro+ Mattress with Advanced Microclimate® Technology (AMT) uses a non-powered system of intake and output valves to provide weight-based pressure redistribution whenever a patient moves or is repositioned. The Pro+ hospital bed surface is a mattress used with various Hillrom hospital beds such as Centrella, Versa care, Advanta, Careassist and Hillrom900 beds.

A total of 139 complaints from one customer were received for Centrella bed and mattress issues with Low Air Loss (LAL) errors on the mcm display system from 1 Jan 2019 (earliest distribution date) through 1 March 2022. Initially complaints were raised against the beds, however investigation determined the issue was with the Pro+ Mattress blower system.

On 9 March 2022, during complaint investigation, service technicians troubleshooted the issue by reconnecting the system to fix the alarm and error. Log file review later identified several errors in the display which confirmed that the troubleshooting (reconnection of the systems) did not solve the issue. The engineering team determined that the error messages were caused by the blower of the mattress with Low Air Loss (LAL).

The impact of this failure mode can cause a reduction in the microclimate management performance of the Pro+, which could result in an increased risk to skin and result in potential pressure injury. Microclimate management is used to reduce skin temperature and moisture of patient skin.

No adverse patient experiences have been reported.

Recall start date: July 1, 2022

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime