· Published by Health Canada · Official notice
| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| More than 10 numbers, contact manufacturer. | 06917259001 | Ventana He 600 System |
This recall is regarding fluid leaks within the middle stainer module of the Ventana he 600 that likely resulted in electrical shorts at the plate heater connections that then escalated into fires.
Recall start date: June 16, 2022