Product: Ventana HE 600 System
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
More than 10 numbers, contact manufacturer.06917259001Ventana He 600 System

Issue

This recall is regarding fluid leaks within the middle stainer module of the Ventana he 600 that likely resulted in electrical shorts at the plate heater connections that then escalated into fires.

Recall start date: June 16, 2022