Health Products

BD Veritor Plus Analyzer

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: BD Veritor Plus Analyzer
Issue: Medical devices - Performance
What to do: Review/share recall notice, follow recommendations until replacement, return completed response form.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.BD Veritor Plus Analyzer256066

Issue

BD has now received 6 complaints for specific product code 256066 that the product referenced above has the potential to cause fire. BD's initial investigation indicated that the issue occurred simultaneous with the use of an unauthorized alternative AC/DC power adapter, and not the approved power provided by BD with the BD Veritor Plus Analyzer. However, the most recent complaint occurred while using the BD Veritor Plus Analyzer with the BD provided AC/DC power.

Recall start date: July 6, 2022

Get told about the next health product recall.