· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| Not applicable. | Emblem S-Icd System | 3200 |
| Not applicable. | Latitude Programming System | 3300 |
In December 2020, Boston Scientific committed to developing a software enhancement that detects and alerts Healthcare Professionals (HCPs) if an Emblem S-Icd exhibits hydrogen-induced accelerated battery depletion. This enhanced software allows clinicians to identify Emblem S-Icds exhibiting hydrogen-induced accelerated battery depletion sooner. This enhanced software is now available, and sales professionals are upgrading the programmers.
Recall start date: July 7, 2022