Health Products

HLS Set Advanced Bioline Coating

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Gastroenterology and urology
Product: HLS Set Advanced Bioline Coating
Issue: Medical devices - Labelling and packaging
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
3000183179
3000183910
3000183908
3000183172
3000183191
3000184681
HLS Set Advanced Bioline Coating70105-2794 BEQ-HLS 7050-USA
70105.2797 BEQ-HLS 5050-USA

Issue

The former packaging design of HLS set contains accessories packed in medical paper as opposed to the new introduced packaging design in Tyvek. Based on design verification tests it showed that medical paper is a weaker technical solution than Tyvek, carrying a potential risk of breaching and thus compromising the product sterility. Therefore it was decided to remove all HLS sets with former packaging design from the market. New packaging was introduced in September 2021.

Recall start date: July 18, 2022