· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| 3000183179 3000183910 3000183908 3000183172 3000183191 3000184681 | HLS Set Advanced Bioline Coating | 70105-2794 BEQ-HLS 7050-USA 70105.2797 BEQ-HLS 5050-USA |
The former packaging design of HLS set contains accessories packed in medical paper as opposed to the new introduced packaging design in Tyvek. Based on design verification tests it showed that medical paper is a weaker technical solution than Tyvek, carrying a potential risk of breaching and thus compromising the product sterility. Therefore it was decided to remove all HLS sets with former packaging design from the market. New packaging was introduced in September 2021.
Recall start date: July 18, 2022