Health Products

Em2400 Exactamix Valve Set

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Em2400 Exactamix Valve Set
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
Lot 60316024 and higherH938724Em2400 Exactamix Valve Set

Issue

Baxter is issuing an urgent medical device correction communication regarding the potential for leaking valves in the Exactamix 2400 valve sets. Baxter has observed an increase in complaints for leaking of ports 1 and 2 when in the closed position, resulting in unintended ingredient transfer into the compounded admixture. This leak would be most noticeable when the Exactamix compounder is pumping and moving fluid through the common fluid pathway. The affected product has been distributed since 11/28/2021 in Canada.

Recall start date: July 18, 2022