NexSys PCS 300/ PCS2 Plasma Collection System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Gastroenterology and urology, Health products - Medical devices - Haematology
Product: NexSys PCS 300/ PCS2 Plasma Collection System
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
All lots.NexSys PCS 300 Plasma Collection SystemPCS-300-INT
All lots.PCS206002-110-NA-EW
06002-CP-110
06002-110-EXP-EW
06002-CPL-110
06002-110-NA

Issue

A software defect is present where in the case that the customer is running on smart last cycle, and a procedure is aborted or has an unrecoverable interruption, the red blood cell loss estimation flag may not be accurate, and should not be used as the basis for donor deferral in the case of RBC loss.

Recall start date: July 15, 2022