· Published by Health Canada · Official notice
| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| VERSION 3.7.2 AND BELOW | 684 4458 | Vitros Xt 3400 Chemistry System |
| VERSION 3.7.2 AND BELOW | 6844461 | Vitros Xt 7600 Integrated System |
Ortho Clinical Diagnostics (Ortho) became aware of an anomaly that affects Vitros Xt 3400 and Vitros Xt 7600 systems. These systems utilize a digital imaging reflectometer and led light source to generate Vitros microslide assay results. Ortho's on-going investigation has determined that performing a system shutdown on Vitros Xt 3400 and Vitros Xt 7600 systems may, in some instances, cause a positive or negative shift in the electrical current supplied to the led used in conjunction with the digital imaging reflectometer. This shift in electrical current may also lead to a shift in multiple Vitros microslide assay results reported by the system.
Recall start date: July 22, 2022