Health Products

Heat And Moisture Exchanger (HME) And Filters

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Heat And Moisture Exchanger (HME) And Filters
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
KMZ21K0182N/AHeat And Moisture Exchanger (HME) And Filters

Issue

Teleflex is initiating this voluntary recall for the above-mentioned products due to reports of the device splitting or detaching during use. The devices are intended for connection to respiratory equipment in intensive care units and operating theatres to protect the equipment from potential airborne contaminants. If the filter splits or detaches during use, the functionality and efficiency of the device may be compromised, potentially resulting in a leakage and insufficient air supply to the patient with potential desaturation.

As of August 2022, a total of 36 complaints reporting the filter splitting or detaching have been received for the products in scope of this recall. Of these 36 complaints, 4 complaints reported injuries including desaturation. No deaths or long-term patient injuries have been reported at this time.

Recall start date: August 29, 2022