Health Products

Xpert EV Assay

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices
Product: Xpert EV Assay
Issue: Medical devices - Performance
What to do: Discontinue use of affected product, complete response form and dispose of remaining inventory.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
1000408741/22001GXEV-100N-10Xpert EV Assay

Issue

On 27-Jul-2022 a non-conformance investigation was initiated for an issue related to a component of the test. In early August 2022, the correlation was made that this component issue may have resulted in customer complaints of invalid and false negative test results that was being investigated separately. The investigation showed that the beads used in the tests caused low sample processing control/internal control endpoints, which could lead to false negative and invalid test results.

Recall start date: August 25, 2022

Get told about the next health product recall.
Get health product recall alerts