· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| HS00191 HS00241 HS00240 HS00256 HS00242 | Alinity Hs Slide Maker Stainer Module | 09P69-01 |
| HQ00903 HQ00900 HQ00893 HQ00410 HQ00411 HQ00787 HQ00905 HQ00899 HQ00432 HQ00906 HQ00792 HQ00901 | Alinity Hq Analyzer | 09P68-01 |
The Alinity h-series system control center (SCC) application software version 4.3 (and below) and the local user interface (LUI) embedded analyzer engine (EAE) contain a cybersecurity certificate that expired on 08sep2022. Whenever the software checks for the cybersecurity certificate expiration date and it detects that it has expired, the SCC and/or LUI becomes nonfunctional. This renders the Alinity h-series unable to process samples. There is no impact to cybersecurity or results integrity on the Alinity h-series with this issue.
Recall start date: September 12, 2022