Health Products

Leica Microsystems M530 Optics Carrier

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Ear, nose and throat
Product: Leica Microsystems M530 Optics Carrier
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
190122003
191121006
11221001
140222002
60422002
70422001
Leica Microsystems M530 Optics Carrier10448737
10448704

Issue

During internal testing Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier. The photodiodes are used as luxmeters to optimize the "brightcare plus" limit which is calculated based on real-time light intensity data to compensate for decreased light intensity as bulbs age. The component change of the photodiode will result in an inaccurate adjustment of the illumination limits.