Health Products

Artis Pheno System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Artis Pheno System
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
164123
164101
164221
164367
Artis Pheno System10849000

Issue

The Artis Pheno System is provided with a dedicated error management mechanism. In rare cases of specific X-ray tube failures, the system is intended to automatically switch over to "bypass fluoroscopy" mode. However, in case such a failure occurs while the "block radiation" functionality is active, the "block radiation" functionality cannot be unlocked, and the system will not switch over to "bypass fluoroscopy" mode. In "bypass fluoroscopy" mode a limited imaging functionality (nonsubtracted, Continuous fluoroscopy with reduced power and without acquisition and storage of images) would remain available.

Recall start date: September 23, 2022

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime