Health Products

Clearpro Double Swivel Closed Suction Catheter

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Clearpro Double Swivel Closed Suction Catheter
Issue: Medical devices - Performance
What to do: Identify, quarantine and destroy any affected lots.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
Lot 6922090011DYNCPDS10Clearpro Double Swivel Closed Suction Catheter

Issue

Clearpro Closed Suction Catheter can potentially come apart from the connector near the thumb valve. This can happen when the catheter is being pulled out of the artificial airway in the patient suctioning process.

Recall start date: August 9, 2022