Health Products

Mas Omni Cardio

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Mas Omni Cardio
Issue: Medical devices - Labelling and packaging
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
OCRD2503S
OCRD2503L
OCRD2503U
Mas Omni CardioOCRD-SP
OCRD-UL
OCRD-L

Issue

Manufacturer initiated recall notifying users of lot-specific open vial stability claim reduction = IFU correction to affected lots. Open vial stability for troponin I present in OCRD-UL, OCRD-I and OCRD-SP does not meet the 15 day claim in IFU. Analysis has demonstrated that the analyte is stable for 5 days once the vial is opened. Informing cosignees prior to use via each unit including a product bulletin and labelling on orange paper with corrected open vial stability.

Recall start date: September 29, 2022

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime