Health Products

Ortho Vision Analyzers

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Haematology
Product: Ortho Vision Analyzers
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
More than 10 numbers, contact manufacturer.6904577Ortho Vision Analyzer
70002012
70002022
70003019
70002013
70002027
70003018
70003020
70002021
6904576Ortho Vision Max Analyzer

Issue

During normal operation of the ortho vision analyzer, the camera imaging subsystem (CIMS) interfaces with a hardware camera to request and retrieve camera images for column grading and results interpretation. The camera software uses a first-in, first-out queuing system to take and return images to the CIMS software. Under an atypical software anomaly, images used for analysis of a test result did not come from the front and back sides of the same card. Ortho determined that the CIMS processed images out of synchronization (images captured earliest in sequence are placed at the top of the queue to be processed). As a result, image processing software utilized on the ortho vision analyzer may use mismatched front of card and back of card images for column grading and result interpretation.