Health Products

Omnipod Dash System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Omnipod Dash System
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: General public
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.Omnipod Dash, Pdm, Personal Diabetes ManagerPDM-CAN-D001-MM
139723Omnipod Dash, Personal Diabetes Manager KitSKT-CAN-D001-MM

Issue

Through post market surveillance and investigation of returned complaint devices, Insulet has been made aware of pdm battery issues, including battery swelling, fluid leaking from the battery, and in rare cases, extreme overheating, which may pose a fire hazard (thermal runaway events). The health hazard evaluation (HHE) was completed on October 10, 2022 with an analysis of possible causes and hazardous situations with the information that was known at the time.