Health Products

Emblem MRI S-ICD PG

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Emblem MRI S-ICD PG
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
169286Emblem MRIi S-ICD PGA219

Issue

A subset of Emblem S-ICDs was manufactured in September and has an incorrect date/timestamp which may cause an inaccurate display of battery capacity. The device's manufacturing date/timestamp is used by the programmer and latitude™ patient management system to calculate and display the battery capacity. The battery capacity value is intended to decrement in a linear manner over the device's implant life. For devices with an incorrect manufacturing date/timestamp, the initial battery capacity displayed after exiting storage mode is approximately 10%. This displayed battery capacity behavior does not affect the device's performance (i.e., ability to detect/treat arrythmias) or the actual battery longevity.

Recall start date: October 3, 2022.

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime