Health Products

Liberator Stationary Liquid Oxygen Device

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Liberator Stationary Liquid Oxygen Device
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
CBB3022320074
CBB3022310852
CBB3022310843
CBB3022310827
CBB3022310823
CBB3022310842
CBB3022310845
CBB3022320077
CBB3022310792
CBB3022310844
Liberator Stationary Liquid Oxygen Device13261699
13256195

Issue

There were inconsistencies in weld penetration on the longitudinal weld seam on the inner bottle. This can lead to premature inner vessel failure. The failure mode would be potential loss of vacuum, which could cause either an over-delivery of oxygen or the outer vessel to frost. Increased flow of oxygen from the relief valves could lead to an increased fire risk.

Recall start date: October 11, 2022