· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| X206AW8012 X159AW7380 X206AW8010 X206AW8011 X159AW7378 X206AW8020 X159AW7375 | Progressa (Frame) | P7500A000158 P7500A000842 |
During verification testing, it was discovered that a subset of the beds under test were experiencing bent compression links when raising the head section and when maximum patient weight was simulated. This failure mode was due to a change that was made to how the compression links attach to the bed frame. This may cause the compression links to bend under normal use with heavy (500 lbs) patients.
Recall start date: October 25, 2022