Health Products

2.5mm Lifeport Adapters and Patient Circuit Kits

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: 2.5mm Lifeport Adapters and Patient Circuit Kits
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
220423202.5mm Lifeport Adapters 10-Pack9025
22052381
22052407
22062431
22062478
22062458
Patient Circuit Kits938

Issue

Bunnell has received reports of 2.5 mm Lifeport adapters with deformations at the patient end tip. These deformations may result in a narrowing of the effective internal diameter of the Lifeport. An evaluation of returned adapters has confirmed the presence of deformations. Bunnell incorporated is voluntarily recalling a limited number of 2.5mm Lifeport adapters (lot number 22042320) for use with the lifepulse high frequency ventilator. The affected 2.5mm Lifeport adapters were packaged in these products: catalog #9025 (2.5 mm Lifeport adapters 10-pack) and catalog #938 (patient circuit kits).

Recall start date: November 2, 2022

Get told about the next health product recall.
Get health product recall alerts