Health Products

Ortho Optix™ Reader

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Haematology
Product: Ortho Optix™ Reader
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
More than 10 numbers, contact manufacturer.Ortho Optix™ Reader6842222

Issue

Ortho confirmed when multiple orders are processed for the same patient on the Ortho Optix™ Reader, the software may not be able to identify the orders are for the same patient and may not perform the discrepant result checks in 4 different scenarios listed below.

Scenario 1: 1. an lis order record containing 'attending physician' data (field p14) is imported. 2. a manual order is created for the same patient without any modifications to patient demographics (using patient search). 3. the orders are processed to completion. in this scenario, the 'attending physician' information is not included in the manual order. the software detects the manual order as a different version of the patient than the lis order (containing the 'attending physician' data) and the discrepant result check is not performed.

Scenario 2: 1. a manual order is created. 2. an lis order is created for the same patient containing 'attending physician' data. 3. the orders are processed to completion. if these steps occur, the software detects the lis order (containing the 'attending physician' data) as a different version of the patient than the manual order and the discrepant result check is not performed.

Scenario 3: 1. an order is created by either importing an lis order or creating a manual order for a new patient. 2. a manual order is created for the same patient (using patient search) and the patient demographic is modified (for example first name, last name, birthdate, or national id). 3. the orders are processed to completion.

Recall start date: November 10, 2022

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime