Health Products

Irispec Ca/Cb/Cc

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: Irispec Ca/Cb/Cc
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
565-22 and higherIrispec Ca/Cb/Cc800-7211

Issue

Beckman Coulter has become aware of an increase in customer complaints generated for intermittent Irispec Cb glucose control false positive results when used in conjunction with ichemvelocity urine chemistry strips PN 800-7212. If a QC run fails, the system will not allow the operator to run patient samples. the QC run should be repeated and be successful before running patient samples ([ichemvelocity urine chemistry system instructions for use (IFU) P/N 301-7146be, chapter 4; quality control). This issue affects Irispec Cb control lot numbers 565-22 and higher when analyzed with ichemvelocity urine chemistry strips (pn 800-7212) lot numbers 7212554m and higher. There is no impact to patient results, Irispec Ca or Cc controls.

Recall start date: November 14, 2022

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime