· Published by Health Canada · Official notice
| Lot or serial number | Model or catalogue number | Affected products |
|---|---|---|
| More than 10 numbers, contact manufacturer. | MC2-5590S MC2-6090S MC2-4570S MC2-5070S MC2-4590S MC2-5090S MC2-5570S MC2-6070S | Magec 2 (Marketed As Magec X) Device (Various Rod Sizes And Lot Numbers) |
| More than 10 numbers, contact manufacturer. | MS2-6090R MC2-4590R MC2-5070R MC2-4570R MC2-5590R MC2-6070R MC2-5570R MC2-5090R | Magec 2 (Marketed As Magec X) Device (Various Rod Sizes And Lot Numbers) |
Recall relates to a packaging closure integrity issue. Investigation revealed a small population (3.8%) of packaged units have an undersized o-ring gland diameter that can affect the o-ring compression when the hard cap is affixed to the packaging tube. This lack of full compression can create a situation where the integrity of the container closure system is compromised potentially impacting the sterile barrier system.
Recall start date: November 1, 2022