Health Products

BD Alaris Large Volume Pump

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: BD Alaris Large Volume Pump
Issue: Medical devices - Device compatibility
What to do: Review/perform action in recall notice, post recall notice on/near device, complete/return reply form.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
All lots.BD Alaris Large Volume Pump8100

Issue

BD is recalling the product because if a third-party screw is used, or potentially if the BD manufactured pivot latch screw is reused or improperly installed, it may back out, interfering with door closure, which can lead to over or under infusion. The current BD pivot latch screw design, which was introduced in 2009, includes retaining microspheres that prevent the screw from backing out when properly installed in accordance with labeling. We are notifying customers to ensure that (1) all pivot latch screws are inspected by hospital staff prior to use of lvp, (2) all loose, backed out, and third-party screws are replaced by the hospital biomed staff with new BD manufactured pivot latch screws in accordance with the labeling, and (3) customers are aware of the risks associated with backed-out pivot latch screws.

Recall start date: November 15, 2022

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime