Health Products

BBL™ Chocolate II Agar

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: BBL™ Chocolate II Agar
Issue: Medical devices - Performance
What to do: Destroy remaining inventory, share recall notice with users, complete/return response form.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
2206461BBL™ Chocolate II Agar221267

Issue

BD has confirmed through complaint investigation that a single lot of BBL™ Chocolate II Agar has misformulated nicotinamide adenine dinucleotide (NAD) concentration in a portion of the batch. Specifically affected are box numbers 0002 through 0097. These boxes correlate to timestamps on the plates of 11:34 through 13:51. All other box numbers and plate stamps are unaffected and perform as intended.

Recall start date: November 21, 2022

Get told about the next health product recall.