Health Products

Fetal Spiral Electrode

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Obstetrics and gynaecology
Product: Fetal Spiral Electrode
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
Not applicable.Fetal Spiral Electrode9898 031 37631

Issue

Philips has found that the spiral tip of the Fetal Sprial Electrode (FSE) may break off during use, potentially requiring surgical intervention to remove the broken tip from the patient. Based on Philips investigation, this can occur due to over rotation during attachment or pulling the tip from the fetal scalp. The FSE may also break off if the user pulls the spiral tip from the fetal skin, increasing the risk of the spiral tip detaching from the FSE during removal.

Recall start date: November 24, 2022