Heartstart Defibrillation Pads

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Heartstart Defibrillation Pads
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
Not applicableM5071A
M5072A
Heartstart Defibrillation Pads

Issue

Some electrode pads used with the hs1/onsite/home AED have been observed to experience gel separation from the foam/tin backing when peeled from the yellow plastic liner. The gel may fold onto itself, resulting in reduced surface area of gel on the pad. A pad in this condition could cause the hs1/onsite/home AED to deliver less effective or ineffective therapy to the patient due to the reduced surface contact area with the skin.

Recall start date: November 23, 2022