Health Products

Battery Mobile X-Ray Systems Soltus

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Battery Mobile X-Ray Systems Soltus
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
SLS000191
SLS000192
SLS000193
10451Battery Mobile X-Ray System Soltus 450
SLS000131
SLS000130
10501Battery Mobile X-Ray System Soltus 500

Issue

There have been incidents of erratic movement involving Soltus mobile x-ray units. Shis problem is caused by a software problem: the backwards high-speed movement is produced due to a bug in the movement control loop while the bumper is activated, the user pushes "manually" the mobile, and then at a certain moment the bumper is mechanically released. In this situation, the reference values for the movement control loop were not correct, and the unit moved backwards and rotates in high speed.

Recall start date: November 18, 2022

Get told about the next health product recall.
Get health product recall alerts