· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| More than 10 numbers, contact manufacturer. | Heartware Ventricular Assist System - HVAD Implant Kit | 1104 |
Controller driveline cover of HVAD system may become hardened, creating difficulty or an inability to slide the cover back to access the drive line to control the connector. Access to the driveline connector may be delayed, which may result in patient harm from a prolonged pump stop.
Recall start date: November 29, 2022