Health Products

RegenKit® BCT-3

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: RegenKit® BCT-3
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalog number
303
301
308
309
RegenKit® BCT-3BCT-3

Issue

Clients have reported several cases of patients having inflammatory reactions after injection with platelet-rich plasma (PRP), characterized by joint effusion and/or pain. Particles in the sodium citrate solution have been observed, with an irregular appearance of the separation gel and the presence of a white layer on the gel surface. To clarify, the separation gel is used for blood separation. It creates a physical barrier between the red blood cells and the PRP. Only the plasma is reinjected into the patient; the gel is not.

Recall start date: August 2, 2022