· Published by Health Canada · Official notice
| Lot or serial number | Affected products | Model or catalog number |
|---|---|---|
| NMS503505H NMS502986H NMS503341H NMS503484H NMS503162H NMS502821H NMS502807H NMS503187H NMS503195H NMS502816H NMS502805H NMS502845H NMS502840H NMS502849H NMS502814H NMS502820H NMS502356H | Vanta/Sequentia Lt Clinician Programmer Application | A71200 |
In rare occasions, the Vanta patient programmer app (PP APP) a72200 (all versions) can display a "system update needed, service code 303" during the initial interrogation of a Vanta implantable neurostimulator (INS) model 977006. When this code is observed, therapy cannot be adjusted and/or turned off with the PP APP.
Recall start date: November 29, 2022